Insulation Detection and Application of Medical IT Distribution System

<

Xu Shuang

Ankerui Electric Co., Ltd. Shanghai Jiading 201801

Abstract This paper mainly introduces the differences between IEC standards and NSPF standards represented by the United States in the field of medical IT systems, and further understands the different leakage protection technologies of application and medical IT systems.

Keywords leakage current insulation monitoring IEC standard NSPF standard medical IT system

Type 2 medical sites are important places in the hospital, usually operating rooms, intensive care units, or cardiac care rooms. They are places of vital importance to patients and are also places full of medical devices and medical electrical equipment. Leakage protection has special requirements compared to ordinary buildings, that is, through local IT systems. In order to better understand the different leakage protection technologies applied to medical IT systems, first introduce the current relevant international standards.

1 Different standards of medical IT systems in different countries

The main countries and their national standards for adopting medical IT systems in medical type 2 places are as follows:

United States

NFPA 99

Switzerland

MED 4818

United Kingdom

BS7671-SGN7.10 (2003)

Hungary

MSZ2040

Germany

DIN VDE 710

Belgium

TN013

France

NF C15-211

Brazil

NBR13.543

Netherlands

NEN 3134

Chile

N.SEG 4EP79

Italy

TC 10

Canada

CSAZ31.1

Austria

OVE-EN7

Japan

JIS T 1022

Norway

NVE-1991-FEB

Australia

AS3003

Finland

SFS 4372

China

GB 16895.24 - 2005

Spain

UNE20-615-80

International Electrotechnical Commission (IEC)

IEC60364-7-710

Ireland

TC 10

A review of the specifications of the aforementioned countries can be divided into two major standards, namely the standard for monitoring the insulation value of the IT system against the ground, represented by IEC60364-7-710. The main participating countries are mostly EU countries such as the United Kingdom and France. Germany, Finland, Italy, Switzerland, Hungary, Ireland, etc. Outside the European Union countries such as China and Brazil. The other standard, represented by NEPA99 in the United States, monitors the sum of capacitive leakage currents and resistive leakage currents of IT systems. The main countries followed are Canada, Japan, Australia, the Netherlands, Norway, Chile, and many Central American countries.

2. Medical smart insulation monitor

The IEC60364 standard stipulates that for medical electrical equipment used for life-supporting or surgical operations in medical Class 2 sites, equipment with a rated voltage exceeding AC25V or DC60V must use an IT power supply system with an insulation monitor.

Figure 1 Ankerui Medical Intelligent Insulation Monitor

The medical intelligent insulation monitor is installed between the IT system and the earth. When the IT system is grounded, the insulation monitor immediately sends an alarm signal notifying the operator that the system has been running with a fault. The fault must be found as soon as possible so that the second point can be grounded. Troubleshooting before. For the insulation monitor installed in the IT system of the medical type 2 site, in addition to the general standard IEC 61557-8 of the insulation monitor, the special requirements of the IEC 60364-7-710 standard must also be met:

1. Internal internal impedance should be at least 100kΩ;

2. The test voltage should not be greater than DC 25V;

3. Even in the fault conditions, the peak value of the injected current should not exceed 1mA;

4. The signal should be generated at the latest when the insulation resistance drops to 50kΩ, and instruments that test this function should be configured;

5. The Insulation Monitor signals when the ground is interrupted or the electrical line is interrupted.

3. Measuring principle of insulation monitor (IMD)

The measurement principle of the insulation monitor is shown in Figure 1.

Figure 2 Measuring principle of insulation monitor

The AIM-M100 medical insulation monitor is connected to the monitored system by an internal coupling resistor RA. The signal source G inside the insulation monitor sends a measurement voltage signal which is superimposed between the monitored system and the ground (PE line) through the coupling resistance RA, the measurement resistance Rm and the low-pass filter circuit. The measurement voltage passes through the external IT system pair. The total insulation impedance RF of the ground constitutes the measurement loop, so the total insulation resistance of the monitored system to ground is obtained by measuring the measurement current Im in the loop.

The traditional application and measurement of AC system ground insulation resistance insulation monitor often use DC signal measurement method, the relevant circuit diagram can be simplified into Figure 2;

Figure 3 Simplified circuit diagram

The insulation monitor sends a constant current measurement voltage signal. The measurement current Im generated by the insulation resistance of the system and the insulation resistance of the system to ground become linear. The relationship is as follows:

By measuring the measured current Im, the insulation resistance Rf of the system to ground can be easily obtained.

However, this measurement method is easily affected by the DC load current in the monitored system and the AC currents of various frequencies. Therefore, the improved insulation monitor uses the method of superimposing pulse measurement voltage to measure the insulation value of the IT system against the ground. In the last 20 years due to the rapid development of computer technology, ACREL used low-frequency AC signals to measure the insulation resistance of the system. The amplitude and frequency of the measurement signal are controlled by a single-chip microcomputer and can accurately measure the insulation resistance and system capacitance of the current system. Through extensive application in hospitals around the world, it has proven to be a reliable method of measuring insulation on-line.

4. Application Plan of Medical Intelligent Insulation Monitor in Medical Class 2 Locations

The scheme of applying the medical intelligent insulation monitor in the actual field is shown in Figure 4 and Figure 5, and the scheme configuration is shown in Table 1 and Table 2:

Figure 4 Operating room isolated power system solution diagram

Scheme configuration table

Table 2 Isolation Power System Solution Configuration for ICU

5 Quotation plans for main products and power cabinets

In order to facilitate our customers' choice of Acryle products, we now list product quotation, as shown in the following table:

5.1 GGF-I Series ICU/CCU Ward Isolation Power Cabinet

shape

model

specification

unit

Price (ten thousand yuan)

GGF-I3.15

3.15KVA

set

3.8~5.3

GGF-I6.3

6.3KVA

set

4.4~5.9

GGF-I8

8KVA

set

5~7

GGF-I10

10KVA

set

6.8~9.8

ups

Optional, additional price

Dual power switching device

Optional, additional price

Freight

Reference logistics company quote

Note

The above quotation includes commissioning fees, excluding freight and installation fees

5.2 GGF-O Series Operating Room Isolated Power Supply Cabinet

shape

model

specification

unit

Price (ten thousand yuan)

GGF-O6.3

6.3KVA

set

6.4~7.9*

GGF-O8

8KVA

set

7~9*

GGF-O10

10KVA

set

8.6~11.6*

ups

Optional, additional price

Dual power switching device

Optional, additional price

Freight

Reference logistics company quote

Note

The above quotation includes commissioning fees, excluding freight and installation fees

5.3 Medical IT Insulation Monitoring Products

shape

name

model

Features

price

Optional and price (yuan)

Insulation monitor

AIM-M100

IT system ground insulation resistance, load current, transformer temperature monitoring

15600

Dual communication +800 Fault alarm output +400 Temperature detection +800 Load detection +800 Centralized monitoring software +3000 Disconnection detection +800 SOE event record +2000 Self-test function +1000 Support fault location function (AIM-M200 support only) + 2000

AIM-M200

19600

DC power supply

ACLP10-24

Instrument dedicated power supply

300

/

Alarm and display

AID100

Insulation resistance, transformer load rate; remotely set the insulation monitor alarm threshold

/

AID120

/

AID200

Using CAN communication, can remotely display and set the parameters of up to 16 sets of insulation monitoring system, real-time monitoring whether the communication is in good condition

900

/

Current Transformer

AKH0.66-P26

Real-time measurement of transformer load current

150

/

Test signal generator

ASG100

Using CAN bus technology, data exchange with other devices can occur. When an insulation fault occurs in the system, a fault location signal can be injected into the system to achieve L1, L2 disconnection detection.

3780

/

Fault location (evaluation) instrument

AIL100-4

Adopt CAN bus technology, when the insulation fault occurs in the system, it can cooperate with ASG100 to realize fault location function

6050

/

AIL100-8

8050

/

6. Some optimization measures in practical application
1. There is no strict regulation on the placement of isolation transformers for medical IT systems. The location of the isolation transformer determines the distance from the transformer to the system load (medical electrical equipment). The longer the distance, the greater the capacitive current generated by the circuit. The U.S. standard IT system isolation transformer is usually mounted in the wall of the operating room to reduce the distance from the transformer to the medical electrical equipment. To meet Canada's stricter regulations, the isolation transformer of the operating room is suspended on the operating bed like a pylon. Its purpose is also to reduce the capacitive leakage current. In the standards of Australia and Japan, it is stipulated that the cable from the transformer to the operating room does not allow the use of metal conduit, but also for the same purpose. Regardless of whether the transformer is placed inside the operating room, whether it is embedded or suspended, it has hindered the requirement of air cleanliness of the operating room; at the same time, the capacity of the transformer cannot be too large, which also limits the flexibility of the operating room function. If the plastic threading tube is placed in the ceiling of the operating room, the reliability of the electrical installation will be reduced. The IEC specification stipulates that the IT system isolation transformer can be located outside the operating room, which greatly reduces the difficulty of electrical installation, and at the same time it is also very beneficial to the heat dissipation of the transformer.

2. In order to both take into account the practicality of IEC specifications and to minimize capacitive currents in IT systems, the IEC 60364-7-710 standard strictly stipulates that the capacity of isolation transformers must not exceed 10 kVA, which can prevent to a certain extent Medical IT systems that use very long lines.

3. IEC standard isolation transformers have a relatively low inrush current, which reduces the need for an automatic power switching device for the IT system's front stage. The IEC 61558-2-15 standard specifies the isolation transformers. The isolation transformers complying with this specification have relatively small volume and low noise level. The inrush current is less than 12 times the rated current, and the small inrush current is undoubtedly the opposite. The requirements of the UPS connected to the primary of the isolation transformer are reduced. In order to meet the low level ground leakage current requirements of the US standard isolation transformer, isolation transformers often have high noise, and the inrush current is often about 20 times of the rated current. In practice, a set of UPS devices is often connected to multiple isolation transformers. The higher the inrush current of the isolation transformer, the higher the cost of the UPS backup power supply.

7. Conclusion

In summary, in the field of medical IT systems, IEC specifications have strong practicality and are widely accepted and adopted worldwide. The NSPF standard represented by the United States is more stringent than the IEC standard, and the difficulty and cost of implementing the standard are higher than the IEC standard.

references:

[1]Leakage detection and application of medical IT system. Intelligent building electrical technology 2010.3

[2] Ankerui Electric Co., Ltd. Product Selection Manual 2013.1.

About the Author:

Xu Shuang, female, undergraduate, Ankerui Electric Co., Ltd., the main research direction for the smart grid power distribution, Email: Mobile QQ

Curved Bar

Curved Bar,Curved Wet Bar,Bar Curved,Curved Corner Bar

Nantong Dongfang fitness equipment Co., Ltd , https://www.dongfangfitness.com